1/2ページ
ダウンロード(388.5Kb)
Real time measurement of leakage rate and internal pressure
of various masks at the same time
このカタログについて
| ドキュメント名 | Real time mask fitting tester MT-R1 |
|---|---|
| ドキュメント種別 | 製品カタログ |
| ファイルサイズ | 388.5Kb |
| 取り扱い企業 | 柴田科学株式会社 (この企業の取り扱いカタログ一覧) |
この企業の関連カタログ
このカタログの内容
Page1
MT-R1_en_表
Real time mask fitting tester MT-R1
Real time measurement of leakage rate and internal pressure
of various masks at the same time
MT-R1 can measure the adhesiveness of respiratory protection with electric fan or dust mask used in various worksites
continuously and in real time.
As measurement result is displayed graph in the main unit, you can confirm the correct wearing of mask checking the
fitting status.
MT-R1型
Feature
●By using indoor dust, you can measure continuously leakage
rate and internal pressure of mask at wearing various masks
●Measurement result is displayed graph and numeric value in
the main unit and updated in real time.
●Easy change of setting by touch panel in the display.
●After measurement of particles inside and outside of mask
for certain period, it can display the measurement result of
leakage rate.
●By inserting the experimental guide bar into the mask,
automatically it will start measurement.
●Diameter of measured particle can be chosen at either more
than 0.3 μm or more than 0.5μm.
●Special case is included.
■Use
●Metal welding site, Nanomaterial, Road, Asbestos, Exposure
prevention for the people working in nuclear facility
(Confirming adhesiveness of mask, training of wearing mask,
education)
●Evaluation of respiratory protection with electric fan
●Confirmation of protection factor
起動画面 漏れ率測定モード
Page2
MT-R1_en_裏
Specification
Model MT-R1
Measurement Leakage rate : 2 units of particle co unter using light scattering method
Method Tetrahedral internal pressure : Silicon diaphragm sensor
Diameter of
Measured Particle Can choose (1) more than 0.3μm or (2) more than 0.5 μm
Measurement Leakage rate : 0~100%
Range Tetrahedral internal pressure : +-500Pa
Graph Dis play of Leakage rate: Update the value calculated from number ratio for minimum 1 sec in every 0.1 secMeasurfment
Result Tetrahedral internal pressure : Update the garaph in every 0.1 sec
Suction Flow Rate Can set in every 1 sec from 1~999 second
1L/min
Leakage Rate
Calculation Formula Leakage rate (%)= Number of particle inside of mask/number of particle outside of mask x100
Usage Environment 0~40 ℃ 30~90%rh (No condensation)
Power AC100V 50/60Hz 2A
Measurement , 430(W)×390(D)×200(H)㎜
Weight Approx. 12kg
Price¥ Please ask
■Consumable Spares
Item
Special Guide 10 pcs/set
Tube 20m
Connector
Zero Filter
Specifications, and appearance described in this document are based on information as of August 10, 2015. They are subject to change without notice for improvement of the product.
1-1-62, Nakane Soka-City, Saitama, Japan
TEL:+81-48-933-1582 FAX:+81-48-933-1591
E-mail:overseas@sibata.co.jp
http://www.sibata.co.jp/english/
201508K